
1/03 UDC 6300 Process Controller Product Manual 233
6.10 Input #5 - Pulse Input Calibration Procedure
Procedure
Follow the procedure in Table 6-8 to calibrate Input 5 Pulse Inputs.
Table 6-8 Input #5 - Pulse Input Calibration Procedure
Step Description Press Action
1
Enter Calibration Mode
SET
UP
until you see
INPUT 5
Lower Display
CALIB
Upper Displa
FUNC
L1/L2
You will see:
CAL IN5
Lower Display
DISABL
Upper Displa
The calibration sequence is enabled and you will
see:
CAL IN5
Lower Display
BEGIN
Upper Displa
At the completion of the sequence, the selection
automatically reverts to disable.
2
Calibrate Zero
FUNC
L1/L2
You will see:
IN5 ZERO
Lower Display
(Value)
Upper Display
The value displayed represents 1/10 of the actual
calibration value.
For example: If 500 Hz is the desired zero value,
enter 50 for input 5 Zero Value.
Minimum Zero Value = 0 Hz
or
to adjust the zero value then go to the next step.
Table 6-8 continued on next page
Comentarios a estos manuales